Thermometers COMMETER -200 to +1700°C for thermocouples K, J, S
Instruments are designed for direct measurement of temperature.
Advantages:
- included traceable calibration certificate
- selection of thermocouple type from instrument keyboard
- automatic compensation of cold junction temperature
- dual line LCD display with special symbols
- adjustable dual alarm with audio indication for each channel
- memory of minimum and maximum values
- function Hold - manual storing of actual values for later displaying
COMMON TECHNICAL PARAMETERS | |
---|---|
Operating temperature range: | -10 to +60°C |
Connector for probe connection: | subminiature connector for thermocouples |
Accuracy without probes - thermocouples K (NiCr-Ni): | ±(0.4°C + 0.1% from reading) from -200 to +1300°C |
Accuracy without probes - thermocouples J (Fe-Co): | ±(0.4°C + 0.1% from reading) from -200 to +750°C |
Accuracy without probes - thermocouples S (PtRh-Pt): | ±(0.85°C + 0.1% from reading) from -50 to +1700°C |
Resolution: |
J,K thermocouples: 0.1°C from -99.9 to +999.9°C, otherwise 1°C S thermocouples: 0.4°C from -50.0 to +999.9°C, otherwise 1°C |
Power: | 9V battery |
Battery life: | typically 2 months |
Dimensions without probes: | 141 x 71 x 27mm |
Warranty: | two years |
Included accessories
Traceable calibration certificate from the manufacturer, instruction manual.
Calibration certificate with declared metrological traceability of etalons is based on requirements of EN ISO/IEC 17025 standard.
Example of calibration certificate.
Included is also plastic case, battery 9V, Instruments are
equipped with connector for external ac/dc adapter.
OPTIONAL ACCESSORIES | ||
---|---|---|
probes | thermocouple probes type "K" NiCr-Ni | |
A1515 | adapter 230V-50Hz/12Vdc. 9V battery should be replaced with NiMH accumulator 9V. | |
A3400 | NiMH accumulator 9V |